OverviewFull-function keypad. Output to safe area printer or software. The 8200IS Series is an advanced indicator with checkweigh features designed for use in hazardous locations
|
Full-function keypad. Output to safe area printer or software. The 8200IS Series is an advanced indicator with checkweigh features designed for use in hazardous locations — ideal for manufacturing processes where equipment is exposed to flammable gases; flammable or combustible liquid-produced vapors; combustible dusts or ignitable fibers; or flyings.
The 8200IS is extremely versatile for a range of environments and weighing demands, from on-site testing where portable battery-powered equipment is needed to large manufacturing facilities with permanent AC installation. Specific applications and environments include fuel refineries, machinery maintenance, chemical plants, chemical storage, distilleries, and textile manufacturing.
8200IS OPERATIONAL FEATURES INCLUDE
8200IS Intrinsically Safe indicator for use in hazardous locations
8BIS Intrinsically Safe battery pack for use in hazardous locations
8CHG Safe area Battery Charger Unclassified, non-hazardous locations only (Ord Loc)
8AIS Intrinsically Safe AC Input Rating
Intrinsically Safe DC Output Rating
Model | 8100IS | 8200IS |
---|---|---|
Warranty | Two Year | |
Power | 8AIS AC Power Supply: 100VAC 50/60 Hz Battery | |
Battery Power | 8BIS: 8 hours of continuous use with 36 hour charge. (Requires 8CIS Battery Charger) | |
Resolution Range | 200 to 100,000d | |
Number of Load Cells | Up to 4, 350 Ohm | |
Load Cell Entity Parameters | Uo 7.14 V, Io 0.7076 A, Po 0.895 W, Co 10.8 uF, Lo 71 uH | |
Controls | Tare, Gross / Net, Zero, Print, Units | Tare, Gross / Net, ZERO, Print, Units, Full keypad with decimal point, Enter, Accum, Product ID, Clear, Setpoint, F1, F2 |
Digital I/O | Two Remote Switch Inputs | |
Temperature Range | 14° to 140° F (-10° to 40° C) | |
8CHG only | 14 F to 86 F (-10 C to +30 C) | |
Analog Sensitivity | 0.16 V/e minimum, 0.5 V/e typical | |
System Linearity: | 0.01% full scale | |
Analog Signal Range | -0.5mV/V to 5 mV/V with 4 and 6 wire input | |
Excitation Voltage | 5 VDC |
|